Agilent Technologies E1401-90493 Manuel d'utilisateur Page 330

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 642
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 329
330 Chapter 2
Performance Verification Tests
39. Residual Responses
39. Residual Responses
The analyzer input is terminated and the analyzer is swept from
150 kHz to 1 MHz. Then the analyzer is swept in incremental 10 MHz
spans from 1 MHz to the upper frequency range. Any responses above
the specification are noted.
There are no related adjustment procedures for this performance test.
Equipment Required
Termination, 50 Type-N (m)
Additional Equipment for 75 Input
Termination, 75 , BNC (m)
Adapter, Type-N (f) to BNC (m), 75
Additional Equipment for Option BAB
Adapter, Type-N (f) to APC 3.5 (f)
CAUTION Use only 75 cables, connectors, or adapters on instruments with 75
input, or damage to the input connector will occur.
Procedure
150 kHz to 1 MHz
1. Connect the 50
termination to the analyzer input as shown in
Figure 2-57.
75
Input: Use the adapter to connect the 75 termination, and
continue with step 4.
Vue de la page 329
1 2 ... 325 326 327 328 329 330 331 332 333 334 335 ... 641 642

Commentaires sur ces manuels

Pas de commentaire