Chapter 2 333
Performance Verification Tests
40. Fast Time Domain Amplitude Accuracy: Agilent E4401B, E4402B, E4404B, E4405B,
and E4407B (Option AYX)
40. Fast Time Domain Amplitude Accuracy:
Agilent E4401B, E4402B, E4404B, E4405B,
and E4407B (Option AYX)
The analyzer amplitude reference signal is used to compare the
amplitude level of a normal sweep time (
5 ms) to a fast sweep time
(
5 ms) using the marker functions. The difference should be less than
the marker readout resolution specification for the fast sweep times.
Equipment Required
Cable, BNC, 122 cm (48 in)
Adapter, Type-N (m) to BNC (f)
Procedure
Fast Sweep Time Amplitude Accuracy
1. Connect the equipment as shown in Figure 2-58.
Figure 2-58 Fast Time Domain Amplitude Accuracy Test Setup
NOTE No test setup is required for the Agilent E4401B.
2. Press
Preset on the analyzer. Press the Factory Preset softkey, if it is
displayed.
3. Set the analyzer as follows:
FREQUENCY, 50 MHz
SPAN
, Zero Span
Sweep
5 ms
Input/Output
(or Input), Amptd Ref (On) (Agilent E4401B only)
Vue de la page 332