Agilent Technologies E1401-90493 Manuel d'utilisateur Page 341

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 642
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 340
Chapter 2 341
Performance Verification Tests
42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B,
E4404B, E4405B, E4407B, and E4408B (Option 1DN)
Setup
4. Read the power level displayed on the power meter, add 20 dB, and
record the result as Test Record entry 1 of the performance
verification test record as the Absolute Amplitude Accuracy.
Absolute Amplitude Accuracy at 50 MHz = ___________dB
5. Set the power meter to dB relative mode as described in the power
meter operation manual so that the readout is in power level relative
to the power level at 50 MHz (press
Rel/Offset, Rel).
6. Press
Single on the analyzer. At each setting, record the power level
displayed on the power meter in Table 2-93.
7. Set the source amplitude to the settings indicated in Table 2-93.
8. Calculate the Vernier Accuracy by adding 4 dB to the Source Vernier
Table 2-93 Vernier and Power Sweep Accuracy Worksheet
Source
Amplitude
Setting
(dBm)
Source Vernier
Setting
(dBm)
Measured
Power Level
(dB)
Vernier
Accuracy
(dB)
18 -2 2)
-19 -3 3)
-20 (Ref) -4 N/A N/A
-21 -5 4)
-22 -6 5)
-23 -7 6)
-24 -8 7)
-25 -9 8)
-26 -10 9)
Vue de la page 340
1 2 ... 336 337 338 339 340 341 342 343 344 345 346 ... 641 642

Commentaires sur ces manuels

Pas de commentaire