Performance characteristics
Acquisition
Samplerate DSO501xA/503xA:2GSa/aeachchannel
DSO505xA:4GSa/shalfchannel
1
, 2 GSa/s each channel
Equivalent-timesamplerate 400GSa/s(whenreal-timemodeisturnedoff)
Memorydepth Standard8Mptshalfchannel1,4Mptseachchannel
Verticalresolution 8bits,upto12inhigh-resolutionoraveragingmodes
Peakdetection DSO501xA:1nspeakdetect
DSO503xA:500pspeakdetect
DSO505xA:250pspeakdetect
Averaging Selectablefrom2,4,8,16,32,64…to65536
Highresolutionmode Averagemodewith#avg=1
12bitsofresolutionwhen≥10µs/div,at4GSa/sor≥20µs/div,at2GSa/s
Filter Sinx/xinterpolation(singleshotBW=samplerate/4orbandwidthofoscilloscope,whicheverisless)
withvectorsonandinreal-timemode
Vertical system
Scopechannels DSO50x2A:2Ch1and2simultaneousacquisition
DSO50x4A:Ch1,2,3,and4simultaneousacquisition
Bandwidth(–3dB)
2
DSO501xA:DCto100MHz
DSO503xA:DCto300MHz
DSO505xA:DCto500MHz
ACcoupled DSO501xA:3.5Hzto100MHz
DSO503xA:3.5Hzto300MHz
DSO505xA:3.5Hzto500MHz
Calculatedrisetime DSO501xA:3.5nsec
(=0.35/bandwidth) DSO503xA:1.17nsec
DSO505xA:700psec
Single-shotbandwidth DSO501xA:100MHz
DSO503xA:300MHz
DSO505xA:500MHz
1 Halfchanneliswhenonlyonechannelofchannelpair1-2isturnedon,oronechannelofchannelpair3-4isturnedon.
2 Denoteswarrantedspecifications,allothersaretypical.Specificationsarevalidaftera30-minutewarm-upperiodand±10°Cfromfirmwarecalibrationtemperature.
13