Agilent Technologies InfiniiVision 5000 Series Guide de l'utilisateur Page 112

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 296
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 111
112 5000 Series Oscilloscope User’s Guide
3 Triggering the Oscilloscope
To use I
2
C triggering
An I
2
C (Inter-IC bus) trigger setup consists of connecting the
oscilloscope to the serial data (SDA) line and the serial clock
(SCL) line, then triggering on a stop/start condition, a restart, a
missing acknowledge, an EEPROM data read, or on a read/write
frame with a specific device address and data value.
1 Press the More key in the Trigger section of the front panel,
rotate the Entry knob until I
2
C is displayed in the Trigger
softkey, then press the Settings softkey to display the I
2
C
trigger menu.
NOTE
For I
2
C decode setup see page 217.
Signals
menu
Trigger on:
condition
Address
Data 2
value
Trigger level
or threshold
Currently selected Clock
or Data channel
I
2
C trigger
Return to
previous menu
Data
value
Vue de la page 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 295 296

Commentaires sur ces manuels

Pas de commentaire