Agilent Technologies InfiniiVision 5000 Series Guide de l'utilisateur Page 137

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 296
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 136
Triggering the Oscilloscope 3
5000 Series Oscilloscope User’s Guide 137
To trigger on a specific field of the video signal
To examine the components of a video signal, trigger on either
Field 1 or Field 2 (available for interleaved standards). When a
specific field is selected, the oscilloscope triggers on the rising
edge of the first serration pulse in the vertical sync interval in
the specified field (1 or 2).
1 Press the Trigger More key, then press the TV softkey.
2 Press the Settings softkey, then press the Standard softkey to
select the appropriate TV standard.
3 Press the Mode softkey and select Field1 or Field2.
Figure 10 Triggering on Field 1
Vue de la page 136
1 2 ... 132 133 134 135 136 137 138 139 140 141 142 ... 295 296

Commentaires sur ces manuels

Pas de commentaire