114 5000 Series Oscilloscope User’s Guide
3 Triggering the Oscilloscope
Data needs to be stable during the whole high clock cycle or
it will be interpreted as a start or stop condition (data
transitioning while the clock is high).
6 Return to the previous menu; then, press the Trigger: softkey
to select one of the following I
2
C trigger conditions:
Start Condition — The oscilloscope triggers when SDA data
transitions from high to low while the SCL clock is high.
For triggering purposes (including frame triggers), a
restart is treated as a start condition.
Stop Condition — The oscilloscope triggers when data (SDA)
transitions from low to high while the clock (SCL) is high.
Missing Acknowledge — The oscilloscope triggers when SDA
data is high during any Ack SCL clock bit.
Address with no Ack — The oscilloscope triggers when the
acknowledge of the selected address field is false. The R/W
bit is ignored.
Restart — The oscilloscope triggers when another start
condition occurs before a stop condition.
SDA
SCL
Address
R/
Ack Data Ack
Stop
Condition
Start
Condition