4 Triggering
144 InfiniiVision Oscilloscopes User’s Guide
Sequence Trigger
Sequence trigger lets you trigger the oscilloscope after finding a sequence of events.
Defining a sequence trigger requires three steps:
1 Define an event to find before you search for the trigger.
The “find” event can be a pattern, an edge from a single channel, or the combination
of a pattern and a channel edge.
2 Define the trigger event.
The “trigger on” event can be a pattern, an edge from a single channel, the
combination of a pattern and a channel edge, or the nth occurrence of an edge from a
single channel.
3 Set an optional reset event.
If you choose to define a “reset” event, the event can be a pattern, an edge from a
single channel, the combination of a pattern and a channel edge, or a timeout value.
Figure 1 Sequence Trigger Flowchart
To access the sequence trigger settings, press the [More] key in the Trigger section of
the front panel, rotate the Entry knob until Sequence is displayed in the Trigger softkey,
then press the Settings softkey to display Sequence Trigger Menu.
Trigger
holdoff
Scope
triggers
Start
Find:?
No
Yes
No
No
Yes
Yes
Trigger on:?
Reset on:?