Serial Decode/Lister 10
InfiniiVision Oscilloscopes User’s Guide 301
c In the Serial Decode Menu, press the Mode softkey.
d Use the Entry knob (or repeatedly press and release the Mode softkey) to select the
FlexRay serial decode mode.
4 Ensure Decode is switched on. The VPT1000 must be connected and running in
asynchronous or synchronous mode for a meaningful display.
5 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data.
6 If decoded data does not appear, press the Settings softkey, then repeatedly press the
Mode softkey (in the VPT1000 Menu) to cycle through the modes and establish
synchronization.
You can use the Zoom horizontal sweep mode for easier navigation of the decoded data.
Display
decode
Decode
mode
VPT1000
menu
Return to
previous menu
Reset
counters