Chapter 4: Testing Performance
DC Gain Accuracy Test
59
DC Gain Accuracy Test
CAUTION Ensure that the input voltage to the oscilloscope never exceeds ±5 V.
Specifications
Equipment Required
Let the oscilloscope warm up before testing
The oscilloscope under test and other test equipment must be warmed up (with the oscilloscope
application running) for at least 30 minutes prior to the start of any performance test.
DC Gain Accuracy ±2% of full scale at full resolution channel scale (±5
o
C from calibration
temperature)
50 W Input Impedance: Full scale is defined as 8 vertical divisions. Magnification is used
below 10 mV/div. Below 10 mV full scale is defined as 80 mV. The major scale settings
are 5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV and 1 V.
1 MOhm Input Impedance: Full scale is defined as 8 vertical divisions. The major scale
settings are 5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, and 5 V.
Description Critical Specifications Recommended
Model/
Part Numbers
Digital Multimeter DC voltage measurement accuracy better than ±0.1% of
reading
Agilent 34401A or
Agilent 3458A
Cable Assembly
(2 required)
50 characteristic impedance, BNC (m) connectors Agilent 8120-1840
DC Power Supply 100 V Agilent 6614C
Adapter BNC Tee (m)(f)(f) Agilent 1250-0781
2 Adapters 2 BNC (f) to dual banana Agilent 1251-2277