Chapter 4: Testing Performance
DC Gain Accuracy Test
63
record the DMM voltage reading (Vdmm- column) and the scope V avg reading (Vscope-
column) in the DC Gain Test (50 W) section of the Performance Test Record.
8 Repeat step 7 for the remaining vertical sensitivities for channel 1 shown in the DC Gain Test
(50 W) section of the Performance Test Record.
9 Press Default Setup, then turn off channel 1 and turn channel 2 display on.
10 Set the DC Supply voltage (External Supply Setting) to +15 mV.
11 Configure the scope to measure V avg on Channel 2.
a Pull down the Setup menu and select Acquisition. When the Acquisition Setup window is
displayed, enable averaging and set the # of averages to 256.
b Change the vertical sensitivity of channel 2 to 5 mV/div.
c Click the V avg measurement icon on the left side of the screen.
d When the Enter Measurement Info window is displayed, ensure that the Vavg function is set up
as follows and then click OK:
Source = Channel 2
Measurement area = Entire Display
12
Press the Clear Display key on the scope, wait for the #Avgs value to return to 256 and then
record the DMM voltage reading (Vdmm+) and the scope V avg reading (Vscope+) in the DC
Gain Test (50 W) section of the Performance Test Record.
13 Repeat step 12 for the remaining vertical sensitivities for channel 2 in the DC Gain Test (50
W) section of the Performance Test Record.
14 Repeat steps 9 through 13 for channels 3 and 4.
15 Repeat steps 1 through 14 using 1 MW for the scope input impedance rather than 50 W. The
steps should be followed exactly except in step 4(c), you should select the 1 MW radio button
from the Impedance section of the Channel 1 Setup dialog box rather than 50 W and all values
should be entered in the DC Gain Test (1 MW) section rather than the DC Gain Test (50 W)
section.