Agilent Technologies Ininiium 90000 Spécifications Page 125

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 148
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 124
Chapter 5: Replacing Assemblies
To remove and replace the USB or GPIB port
125
To remove and replace the USB or GPIB port
Use this procedure to remove and replace the USB or GPIB port (GPIB is an option and may not
be on your oscilloscope). The procedure is exactly the same for both of these ports so only one
will be shown here. When necessary, refer to other removal procedures. The graphics in this
chapter are representative of the oscilloscope at the time of this printing. Your unit may look
different.
1
Disconnect the power cable and remove the cover and top plate.
2
Remove the screw from top edge of chassis (Figure 5-56).
Figure 5-56
3
Undo the connection to the motherboard by disconnecting the cable.
4
Pull the port out of the instrument.
5
To replace one of these ports, reverse the directions.
Vue de la page 124
1 2 ... 120 121 122 123 124 125 126 127 128 129 130 ... 147 148

Commentaires sur ces manuels

Pas de commentaire