126 L4400 User’s Guide
5 L4433A Dual/Quad 4x8 Reed Matrix
34933T-002 Terminal Block for One-Wire Mode
This terminal block with screw- type connections is labeled with the
model number and the abbreviated module name.
All modules that connect to the analog bus are interlock
protected. This means that when an installed module is exposed
(no terminal block or cable is connected), the analog bus relays
and current channels are open and disconnected from the analog
buses. See
page 112 for further information.
If you are using an Agilent terminal block to connect your DUT
to this module be sure to use the 34933T-002 terminal block that
corresponds to the 1-wire configuration mode. Note that an error
will not be generated if you have installed a terminal block that
doesn't match the present module configuration.
NOTE: Analog
Bus connections
are on Matrix 3
and Matrix 4.
Warning -the insulation of the wiring used
with the terminal block must be rated for
the highest voltage that will be present on
the terminal block or on the analog bus.
NOTE
NOTE