44 L4400 User’s Guide
2 Software Installation and Configuration
Figure 2-24. Instrument Firmware Update Complete.
5. Once the carrier update AND instr
ument sub- assembly update (if one
occurs) complete and no instrument activity is indicated by the front
panel LEDs, cycle power on the instrument. Once the power- on sequence
completes, select ‘Refresh All’ in the Agilent Connection Expert (Figure
2- 10).
Agilent Connection Expert may report t
hat the instrument’s configuration
has changed. This is represented by a yellow triangle and an exclamation
point (!) next to the updated instrument. Select (highlight) the instrument
name. Select ‘Change Properties...’ and then click either ‘Test Connection’
or ‘Identify Instrument’ to update Agilent Connection Expert and then
click ‘OK’. Repeat for each updated instrument on the LAN and GPIB
interfaces.