Test System Overview
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Test System Overview
The Agilent Technologies test system is designed to verify the performance of the PSA
series spectrum analyzers and ensure that they meet published specifications. Follow the
links below to view a listing of the test system requirements, test equipment required, a
full description of each performance test, a full description of each adjustment, the
associated troubleshooting procedures should any of the tests fail, and the Test
Management Environment (TME) software help.
Getting Started -- steps for installing and running the test software
Required Equipment -- test equipment, ESD precautions
Performance Tests -- a list of all performance tests
Adjustments -- a list of all adjustments
Troubleshooting -- troubleshooting performance test failures
Memory Initialization Utilities -- utilities to write serial number information, and to reset
instrument statistical data
Test Management Environment (TME) help -- information on setting up and using the
http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/ (1 of 2) [5/29/2003 8:46:47 AM]
Vue de la page 0
1 2 3 4 5 6 ... 132 133

Résumé du contenu

Page 1 - Test and Adjustment Software

Test System Overview Agilent TechnologiesPSA Series Spectrum Analyzers Test and Adjustment Software Test System OverviewThe Agilent Technologies test

Page 2 - Test System Overview

PSA Test Equipment● 0 dB to 40 db Attenuation: ± .005 dB● 41 dB to 80 dB Attenuation: ± .01 dB● 81 dB to 120 dB Attenuation: ± .21 dB4 Mic

Page 3

Frequency Response AdjustmentPower Meter Calibration (Channel A) Power Meter Calibration (Channel B) http://mktwww.soco.agilent.com/field/lp/PSA/PSAca

Page 4

Frequency Response AdjustmentFrequency Response Adjustment (100 kHz to 3 GHz) Splitter Calibration, Step 1Frequency Response Adjustment (100 kHz to 3

Page 5

Frequency Response AdjustmentFrequency Response Adjustment (100 kHz to 3 GHz) Splitter Calibration, Step 2(Preamp on, Opt. 1DS only)Frequency Response

Page 6

Frequency Response AdjustmentFrequency Response Adjustment (100 kHz to 3 GHz) Setup, Step 2http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp

Page 7

Frequency Response AdjustmentFrequency Response Adjustment (100 kHz to 3 GHz) Setup, Step 1(Preamp on, Opt. 1DS only)Frequency Response Adjustment (10

Page 8

Frequency Response AdjustmentFrequency Response Adjustment (< 100 kHz) System CalibrationFrequency Response Adjustment (< 100 kHz) Setuphttp://m

Page 9 - ≤ 1.08:1

Frequency Response AdjustmentFrequency Response Adjustment (Bands 1 through 4) Setuphttp://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Adjus

Page 10 - PSA Test Equipment

Frequency Response Adjustment (Option B7J) Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Response Adjustmen

Page 11

Frequency Response Adjustment (Option B7J)Type N Cable 11500C X XBNC Cable 10503A X XType N (m) to Type N (m) adapter 1250-1475 X XType N (m) to 3.5

Page 12 - PSA Performance Tests

Frequency Response Adjustment (Option B7J) Frequency Response Adjustment (Option B7J) Setup http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhel

Page 13

PSA Performance Tests Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software PSA Series Performance TestsThe Agilent PSA serie

Page 14 - Troubleshooting

Troubleshooting Test Failures Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Troubleshooting Test FailuresIf a test wi

Page 15

Troubleshooting Test Failures1. Residual Responses Troubleshooting2. Displayed Average Noise Level Troubleshooting3. Frequency Reference Accurac

Page 16

Residual Responses Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Residual Responses TroubleshootingRe

Page 17

Displayed Average Noise Level Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Displayed Average Noise L

Page 18

Frequency Reference Accuracy Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Reference Accura

Page 19

Frequency Readout Accuracy Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Readout Accuracy T

Page 20

Count Accuracy Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Count Accuracy TroubleshootingRelated Ad

Page 21

Spurious Responses Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Spurious Responses TroubleshootingRe

Page 22 - Count Accuracy Test Setup

Third Order Intermodulation Distortion Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Third Order Inte

Page 23

Gain Compression Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Gain Compression TroubleshootingRelate

Page 24 - Spurious Responses Test Setup

PSA Performance Tests11. Resolution Bandwidth Switching Uncertainty12. IF Amplitude Ripple (Option B7J)13. IF Phase Ripple (Option B7J)14. Dis

Page 25

Second Harmonic Distortion Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Second Harmonic Distortion T

Page 26

Power Bandwidth Accuracy Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Power Bandwidth Accuracy Troub

Page 27

Resolution Bandwidth Switching Uncertainty Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Resolution B

Page 28

IF Amplitude Ripple Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software IF Amplitude Ripple Troubleshooting

Page 29

IF Phase Ripple Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software IF Phase Ripple TroubleshootingRelated

Page 30

Input Attenuation Switching Uncertainty Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Input Attenuati

Page 31

Display Scale Fidelity Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Display Scale Fidelity Troublesh

Page 32

Absolute Amplitude Accuracy Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Absolute Amplitude Accuracy

Page 33

Noise Sidebands Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Noise Sidebands TroubleshootingRelated

Page 34

Noise Sidebands TroubleshootingOscillator (OCXO) on the A11 Reference Board.Failures at other offsets can be caused by the A12 Synthesizer Assembly. S

Page 35

Residual Responses Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Residual Responses Performance Test

Page 36

Frequency Response Troubleshooting Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Response TroubleshootingRe

Page 37

Frequency Response TroubleshootingIf the frequency response test fails at frequencies below 3 GHz (Low band), the following should be done:Perform the

Page 38

Option 266/Basic Mode Conflicts Agilent TechnologiesPSA Series Spectrum Analyzers Test and Adjustment Software Option 266/Basic Mode ConflictsThe A.03

Page 39

Option 266/Basic Mode Conflictstests/adjustments will have invalid results:● IF Phase Ripple● IF Amplitude Ripple● Absolute Amplitude Accu

Page 40

Residual Responses Performance TestTroubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/resid.htm (2 of 2) [5/29/20

Page 41

Displayed Average Noise Level (DANL) Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Displayed Average

Page 42

Displayed Average Noise Level (DANL) Performance TestTroubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/danl.htm

Page 43 - IF Phase Ripple Test Setup

Frequency Reference Accuracy Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Reference Accur

Page 44

Frequency Reference Accuracy Performance TestFrequency Standard 5071A X XBNC Cable (2 required) 10503A X XFrequency Reference Accuracy Test SetupTro

Page 45

Frequency Readout Accuracy Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Readout Accuracy

Page 46

Test System OverviewTME software NOTE: If your analyzer has firmware revision A.03.xx, click here for information about possible conflicts between Opt

Page 47

Frequency Readout Accuracy Performance TestType N (m) to 3.5 mm (f) adapter 1250-1744 X Frequency Readout Accuracy Test Setup Troubleshooting http://

Page 48

Count Accuracy Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Count Accuracy Performance TestThe test

Page 49 - Measure Source Setup

Count Accuracy Performance TestCount Accuracy Test SetupTroubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/count.

Page 50

Spurious Responses Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Spurious Responses Performance Test

Page 51

Spurious Responses Performance TestPower Sensor 8487A XPower Splitter 11667B X Power Splitter 11667C XAPC 3.5 Cable 8120-4921 X 2.4 mm Cable 8

Page 52

Spurious Responses Performance TestTroubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/spurs.htm (3 of 3) [5/29/20

Page 53

Third Order Intermodulation Distortion Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Third Order Int

Page 54 - Required Test Equipment

Third Order Intermodulation Distortion Performance TestTest Equipment Model NumberE4440A E4443A, E4445AE4446AE4448ASynthesized Sweeper (2 required)836

Page 55

Third Order Intermodulation Distortion Performance Test Third Order Intermodulation Distortion Low Band Test Setup Third Order Intermodulation Distort

Page 56

Third Order Intermodulation Distortion Performance TestTroubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/toi.htm

Page 57

Before You Start Agilent TechnologiesPSA Series Spectrum Analyzers Test and Adjustment Software Before You StartCheck the following before starting th

Page 58

Gain Compression Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Gain Compression Performance TestThe

Page 59

Gain Compression Performance TestDirectional Bridge 86205A X XAPC 3.5 Cable (2 required) 8120-4921 X XBNC Cable 10503A X X2.4 mm (f) to 3.5 mm (f) ad

Page 60

Gain Compression Performance TestGain Compression (7.6 GHz) SetupTroubleshootinghttp://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performanc

Page 61 - (Preamp on, Opt 1DS only)

Gain Compression Performance Test http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/gaincomp.htm (4 of 4) [5/29/2003 8:47:07 AM]

Page 62

Second Harmonic Distortion (SHD) Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Second Harmonic Disto

Page 63

Second Harmonic Distortion (SHD) Performance TestPower Sensor 8481A X XPower Splitter 11667A X X50 MHz Low Pass Filter 0955-0306 X X300 MHz Low Pass

Page 64

Second Harmonic Distortion (SHD) Performance TestTroubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/shd.htm (3 of

Page 65

Power Bandwidth Accuracy Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Power Bandwidth Accuracy Perf

Page 66

Resolution Bandwidth Switching Uncertainty Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Resolution

Page 67

IF Amplitude Ripple (Option B7J) Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software IF Amplitude Ripple (

Page 68

PSA Test Equipment Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Test EquipmentInstrument Critical SpecificationsRecom

Page 69

IF Amplitude Ripple (Option B7J) Performance TestTest Equipment Model NumberE4440A E4443A, E4445AE4446AE4448ASignal Generator E4433B X XPower Meter E

Page 70

IF Amplitude Ripple (Option B7J) Performance TestTroubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/ifamprip.htm

Page 71

IF Phase Ripple (Option B7J) Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software IF Phase Ripple (Option B

Page 72 - PSA Adjustments

IF Phase Ripple (Option B7J) Performance TestType N (f) to BNC (m) adapter 1250-1477 X X2.4 mm (f) to Type N (f) adapter 11903B XIF Phase Ripple

Page 73

Display Scale Fidelity Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Display Scale Fidelity Performa

Page 74 - Lowband Mixer Bias Adjustment

Display Scale Fidelity Performance TestTest Equipment Model NumberE4440A E4443A, E4445AE4446AE4448ADigital Signal Generator 8663A X X1 dB Step Attenu

Page 75

Input Attenuation Switching Uncertainty Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Input Attenuat

Page 76 - SLODA Adjustment Setup

Input Attenuation Switching Uncertainty Performance TestTest Equipment Model NumberE4440A E4443A, E4445AE4446AE4448ASignal Generator 8663A X X1 dB St

Page 77

Absolute Amplitude Accuracy Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Absolute Amplitude Accurac

Page 78 - FELOMA Label Locations

Absolute Amplitude Accuracy Performance TestTest Equipment Model NumberE4440A E4443A, E4445AE4446AE4448ASignal Generator 8663A X XPower Meter E4419B

Page 79

PSA Test EquipmentSweep SignalGenerator(Needed for E4446A, E4448A)Frequency: 10 MHz to 50 GHzFrequency Resolution: 1 HzHarmonic level: < -40 dBcAmp

Page 80

Absolute Amplitude Accuracy Performance Test Absolute Amplitude Accuracy Test Setup Troubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/PSAcal

Page 81 - Second LO Power Adjustment

Noise Sidebands < 50 kHz Offsets Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Noise Sidebands &l

Page 82

Noise Sidebands < 50 kHz Offsets Performance TestTest Equipment Model NumberE4440A E4443A, E4445AE4446AE4448ASignal Generator 8663A X XBNC Cable 1

Page 83

Noise Sidebands > 50 kHz Offsets Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Noise Sidebands &g

Page 84

Noise Sidebands > 50 kHz Offsets Performance TestRequired Test EquipmentTest Equipment Model NumberE4440A E4443A, E4445AE4446AE4448ASignal Generato

Page 85

Frequency Response (Option B7J) Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Response (Op

Page 86 - Attenuator Slope Adjustment

Frequency Response (Option B7J) Performance TestPower Meter E4419A/B X XPower Sensor (2 required) 8482A Option H84 X XPower Splitter 11667A X XType N

Page 87

Frequency Response (Option B7J) Performance TestFrequency Response (Option B7J) Test SetupTroubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/

Page 88

Frequency Response 300 kHz to 3 GHz Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Response

Page 89

Frequency Response 300 kHz to 3 GHz Performance Testwith the preamplifier turned off and 10 dB of RF attenuation. This section will characterize the t

Page 90

PSA Test EquipmentPower Sensor(Needed for E4446A, E4448A)Frequency Range: 50 MHz to 50 GHzAmplitude Range: -20 to +20 dBVSWR: 50 MHz to 100 MHz: 1.15:

Page 91

Frequency Response 300 kHz to 3 GHz Performance Test Power Meter Calibration (Channel B) 300 kHz to 3 GHz Splitter Calibration300 kHz to 3 GHz Flatnes

Page 92

Frequency Response 300 kHz to 3 GHz Performance Test300 kHz to 3 GHz Splitter Calibration(Preamp on, Opt 1DS only)300 kHz to 3 GHz Flatness Test Setup

Page 93

Frequency Response 300 kHz to 3 GHz Performance TestTroubleshooting http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/freqresmid

Page 94 - YTF Alignment Setup

Frequency Response Above 3 GHz Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Response Abov

Page 95

Frequency Response Above 3 GHz Performance TestAbove 3 GHz the flatness specification widens out such that higher uncertainties can be tolerated, ther

Page 96

Frequency Response Above 3 GHz Performance Test Power Meter Calibration (Channel B) Frequency Response Above 3 GHz Test SetupTroubleshooting http://mk

Page 97

Frequency Response Below 300 kHz Performance Test Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Response Be

Page 98 - 10 MHz Reference Connections

Frequency Response Below 300 kHz Performance Testrelative to the amplitude at 300 kHz. This relative flatness is converted to an absolute flatness by

Page 99 - Adaptive Flatness

Frequency Response Below 300 kHz Performance Test Frequency Response Below 300 kHz Test SetupTroubleshooting http://mktwww.soco.agilent.com/field/lp/P

Page 100 - Frequency Response Adjustment

PSA Memory Initialization Utility Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Memory Initialization UtilitiesThe PS

Page 101

PSA Test EquipmentPower Splitter Frequency: DC to 26.5 GHz3.5 mm (f)VSWR: DC to 3 GHz: < 1.06:1 3 GHz to 26.5 GHz: <1.22:1Insertion

Page 102 - Setup, Step 1

PSA Memory Initialization Utility● Reset mechanical attenuator actuations when either the A14 or A15 Input Attenuators are replaced.● Reset 50

Page 103 - Setup, Step 2

PSA Adjustments Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software PSA Series AdjustmentsThe following adjustments can be

Page 104

PSA Adjustments12. Frequency Response Adjustment (Option B7J) Memory Initialization Utilitieshttp://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebh

Page 105 - System Calibration

Lowband Mixer Bias Adjustment Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Lowband Mixer Bias Adjustment The bias cu

Page 106

Lowband Mixer Bias Adjustment http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Adjustment/adjustmixbias.htm (2 of 2) [5/29/2003 8:47:24 AM]

Page 107 - PSA Series Spectrum Analyzers

Switched LO Distribution Amplifier (SLODA) Adjustment (E4440A, E4443A, E4445A) Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment So

Page 108

Switched LO Distribution Amplifier (SLODA) Adjustment (E4440A, E4443A, E4445A)SLODA Adjustment Setuphttp://mktwww.soco.agilent.com/field/lp/PSA/PSAcal

Page 109

Frequency Extended Local Oscillator Multiplier Amplifier (FELOMA) Amplifier Adjustment (E4446A, E4448A) Agilent TechnologiesPSA Series Spectrum Analyz

Page 110

Frequency Extended Local Oscillator Multiplier Amplifier (FELOMA) Amplifier Adjustment (E4446A, E4448A)Required Test EquipmentTest Equipment Model Num

Page 111 - Troubleshooting Test Failures

Second LO Power Adjustment Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Second LO Power Adjustment The 3.6 GHz 2nd L

Page 112

PSA Test EquipmentBNC (m) to SMC (f) 10020-61605 10503A BNC cable with 1250-0832 BNC to SMC adapterAFilters50 MHz Low Pass Cutoff Frequency: 50 MHzRej

Page 113

Second LO Power AdjustmentTest Equipment Model NumberE4440A E4443A, E4445AE4446AE4448APower Meter E4419A/B X XPower Sensor 8481A X XType N (f) to 3.

Page 114

Second LO Power Adjustment http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Adjustment/adjust2lopow.htm (3 of 3) [5/29/2003 8:47:28 AM]

Page 115

Analog IF Input Detector Adjustment Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Analog IF Input Detector Adjustment

Page 116

Analog IF Input Detector AdjustmentPower Meter Calibration Analog IF Input Detector Adjustment Setuphttp://mktwww.soco.agilent.com/field/lp/PSA/PSAcal

Page 117

Attenuator Slope Adjustment Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Attenuator Slope Adjustment The slopes for

Page 118

Attenuator Slope AdjustmentSynthesized Sweeper83630A/B83640A/B, 83650A/B X Synthesized Sweeper 83640A/B, 83650A/B XPower Meter E4419A/B X XPower Se

Page 119

Attenuator Slope Adjustmenthttp://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Adjustment/adjustattslope.htm (3 of 3) [5/29/2003 8:47:30 AM]

Page 120

50 MHz Calibrator Amplitude Adjustment Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software 50 MHz Calibrator Amplitude Adju

Page 121

50 MHz Calibrator Amplitude Adjustmenthttp://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Adjustment/adjustcalpow.htm (2 of 2) [5/29/2003 8:47

Page 122

Overload Detector DAC Adjustment Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Overload Detector DAC AdjustmentThe ov

Page 123

PSA Test EquipmentBNC (m) to SMA (f)Frequency: DC to 1.3 GHzVSWR: ≤ 1.13:11250-1700 PType N (m) to APC 3.5 mm (f)(2 Required, Opt. BAB 3 Required)Freq

Page 124

Overload Detector DAC Adjustmenthttp://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Adjustment/adjustoverdet.htm (2 of 2) [5/29/2003 8:47:32 A

Page 125

10 MHz Internal Frequency Reference Adjustment Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software 10 MHz Internal Frequenc

Page 126

10 MHz Internal Frequency Reference AdjustmentTest Equipment Model NumberE4440A E4443A, E4445AE4446AE4448AFrequency Standard 5071A X X20 dB Fixed Att

Page 127

YTF Alignment (RYTHM and SBTX) Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software YTF Alignment (RYTHM and SBTX)The instru

Page 128

YTF Alignment (RYTHM and SBTX) Test Equipment Model NumberE4440A E4443A, E4445AE4446AE4448ASynthesized Sweeper83630A/B 83640A/B, 83650A/BX Synthesize

Page 129

YTF Alignment (RYTHM and SBTX)http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Adjustment/adjustytf26.htm (3 of 3) [5/29/2003 8:47:34 AM]

Page 130

Frequency Response Adjustment Agilent TechnologiesPSA Series Spectrum AnalyzersTest and Adjustment Software Frequency Response AdjustmentThis adjustm

Page 131 - Related Adjustments

Frequency Response AdjustmentTwo sources, and two sets of power sensors are used to provide the needed frequency range.To measure frequencies below 10

Page 132

Frequency Response AdjustmentPower Meter E4419A/B X XPower Sensor (2 required) 8482A X XPower Sensor 8485A X Power Sensor 8487A XPower Splitter

Page 133

Frequency Response AdjustmentAdaptive FlatnessAdaptive Flatness(Preamp on, Opt. 1DS only)http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/A