Agilent Technologies 83630A Spécifications Page 44

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 133
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 43
Display Scale Fidelity Performance Test
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Display Scale Fidelity Performance Test
This test verifies that the PSA meets its amplitude accuracy specification for amplitude
linearity. The input attenuator is set to 10 dB in zero span mode. The first part of the test
sets the source to 5 dBm and external attenuators are set to 30 dB. The signal is then
measured. The value measured is recorded as the reference level. The RF input is then
varied between the specification intervals of 0 dBm and -17 dBm.
The second part of the test sets the source to -20 dBm and external attenuators are set to 5
dB. The signal is then measured. The value measured is recorded as the reference level.
The RF input is then varied between the specification intervals of -25 dBm and -55 dBm.
Signals below this level are guaranteed by design and are not measured.
NOTE: This test requires step attenuator calibration data to obtain valid results. Make
sure you enter the calibration data for the step attenuators that you're using. Please refer
to the
Enter Equipment Calibration Data section in the TME help for details on how to
enter this data. You must enter the data for the following columns to obtain valid results:
Attn(dB), Uncert(dB) and ReflSize.
Required Test Equipment
http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/scalefid.htm (1 of 2) [5/29/2003 8:47:12 AM]
Vue de la page 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 132 133

Commentaires sur ces manuels

Pas de commentaire