Third Order Intermodulation Distortion Troubleshooting
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Third Order Intermodulation Distortion
Troubleshooting
Related Adjustments
Lowband Mixer Bias Adjustment
Switched LO Distribution Amplifier Adjustment
Second LO Power Adjustment
If TOI fails to meet specifications, suspect the first mixer of the band which fails (A20
Low Band or A19 RYTHM). Next suspect the A10 Third Converter assembly.