Agilent Technologies Agilent 8166A Manuel d'instructions Page 127

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 274
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 126
Signal Generation – The SOURce Subsystem Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 127
command:
[:SOURce[n]][:CHANnel[m]]:POWer:STATe
syntax: [:SOURce[n]][:CHANnel[m]]:POWer:STATe<wsp><boolean>
description: Switches the laser of the chosen source on or off.
parameters: A boolean value: 0: Laser Off
1: Laser On
response: none
example: sour2:pow:stat 1
affects: All laser source, DFB source, and tunable laser modules and return loss modules with an in-
ternal source
command:
[:SOURce[n]][:CHANnel[m]]:POWer:STATe?
syntax: [:SOURce[n]][:CHANnel[m]]:POWer:STATe?
description: Queries the laser state of the chosen source.
parameters: none
response: A boolean value: 0: Laser Off
1: Laser On
example: sour2:pow:stat? 1<END>
affects: All laser source, DFB source, and tunable laser modules and return loss modules with an in-
ternal source
command:
[:SOURce[n]][:CHANnel[m]]:POWer:UNIT
syntax: [:SOURce[n]][:CHANnel[m]]:POWer:UNIT<wsp>DBM|0|Watt|1
description: Sets the power units
parameters: 0 or DBM:
1 or W:
dBm (default)
Watts
response: none
example: sour2:pow:unit w
affects: All tunable laser and DFB source modules
command:
[:SOURce[n]][:CHANnel[m]]:POWer:UNIT?
syntax: [:SOURce[n]][:CHANnel[m]]:POWer:UNIT?
description: Return the current power units
parameters: 0:
1:
dBm
Watts
response: none
example: sour2:pow:unit? +0<END>
affects: All tunable laser and DFB source modules
Vue de la page 126
1 2 ... 122 123 124 125 126 127 128 129 130 131 132 ... 273 274

Commentaires sur ces manuels