Measurement Operations & Settings Triggering - The TRIGger Subsystem
176 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
:TRIGger[n][:CHANnel[m]]:OUTPut
syntax: :TRIGger[n][:CHANnel[m]]:OUTPut
description: Specifies when an output trigger is generated and arms the module.
parameters: DISabled:
AVGover:
MEASure:
MODulation:
STFinished:
SWFinished:
SWSTarted:
Never.
When averaging time period finishes.
When averaging time period begins.
For every leading edge of a digitally-modulated (TTL) signal
When a sweep step finishes.
When sweep cycle finishes.
When a sweep cycle starts.
response: none
example: trig1:outp dis
affects: All tunable laser modules, Agilent 8163A/B series power meters, and Agilent 8161x series
return loss modules.
dual sensors: Can only be sent to master channel, slave channel is also affected.
NOTE In continuous mode, wav:swe:step:[widt] is used for triggering, see page 146.
command:
:TRIGger[n][:CHANnel[m]]:OUTPut?
syntax: :TRIGger[n][:CHANnel[m]]:OUTPut?
description: Returns the condition that causes an output trigger.
parameters: none
response: DISabled:
AVGover:
MEASure:
MODulation:
STFinished:
SWFinished:
SWSTarted:
Never.
When averaging time period finishes.
When averaging time period begins.
For every leading edge of a digitally-modulated (TTL) signal
When a sweep step finishes.
When sweep cycle finishes.
When a sweep cycle starts.
example: trig1:outp? dis<END>
affects: All tunable laser modules, Agilent 8163A/B series power meters, and Agilent 8161x series
return loss modules.
dual sensors: Can only be sent to master channel, slave channel parameters are identical.