Agilent Technologies ENA Series Guide de l'utilisateur Page 301

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 775
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 300
Chapter 7 295
Fixture Simulator
Evaluating Balanced Devices (balance-unbalance conversion function)
7. Fixture Simulator
Evaluating Balanced Devices (balance-unbalance conversion
function)
The balance-unbalance conversion function simulates a measurement under a balanced
state based on measurement results obtained in an unbalanced state (Figure 7-7). This
function enables you to evaluate devices with balanced ports.
NOTE Be sure to set the impedances of the two unbalanced ports equal to each other. For more
details on setting up port impedance for unbalanced ports, refer to “Converting the Port
Impedance of the Measurement Result” on page 288.
Figure 7-7 Balance-unbalance conversion
The types of devices that can be evaluated using the E5070B/E5071B are shown in Figure
7-8.
NOTE To evaluate a balanced device, an E5070B/E5071B with at least three test ports (option
313, 314, 413, or 414) is required.
Vue de la page 300
1 2 ... 296 297 298 299 300 301 302 303 304 305 306 ... 774 775

Commentaires sur ces manuels

Pas de commentaire