486 Chapter 14
Controlling E5091A
Calibration
Calibration
Follow these steps to perform calibration with the E5091A connected:
Step 1. Press (or ) to set the channel that you want to calibrate to the
active channel.
Step 2. Follow “Displaying the E5091A properties” on page 483 to display the E5091A properties.
Step 3. According to the procedures in Chapter 4, “Calibration,” on page 93, perform the
calibration. Check the corresponding E5091A test ports shown in the calibration properties
as the port names of the E5070B/E5071B are displayed on the calibration menu, connect
the calibration standard to the corresponding test ports of the E5091A, and perform the
calibration.