30 7000 Series Oscilloscopes Service Guide
2 Testing Performance
1 Turn on the test equipment and the oscilloscope. Let them
warm up for 30 minutes before starting the test.
2 Set up the oscilloscope calibrator.
a Set the oscilloscope calibrator to provide a DC offset
voltage at the Channel 1 output.
b Use the multimeter to monitor the oscilloscope calibrator
DC output voltage.
3 Use the 8-by-2 test connector and the BNC cable assembly to
connect digital channels D0-D7 to one side of the BNC Tee.
Then connect the D0-D7 ground lead to the ground side of
the 8-by-2 connector. See Figure 2.
Table 4 Equipment Required to Test Digital Channel Threshold
Accuracy
Equipment Critical
Specifications
Recommended Model/Part
Digital Multimeter 0.1 mV resolution,
0.005% accuracy
Agilent 34401A
Oscilloscope Calibrator DC offset voltage
6.3 V
Fluke 5820A
BNC-Banana Cable Agilent 11001-66001 or
Pomona 2BC-BNC-36
BNC Tee Agilent 1250-0781 or
Pomona 3285
50
BNC Cable Agilent 10503A
BNC Test Connector,
8-by-2
User-built (See page 26)
Probe Cable Agilent 01650-61607 or
Agilent 54620-68701
7000 service guide.book Page 30 Thursday, October 16, 2008 12:18 PM