98 7000 Series Oscilloscopes Service Guide
5 Replacing Assemblies
9 Using a TORX T10 remove the four screws that secure the
display to the front deck. The upper right screw also secures
the inverter cable shield. Take care while threading display
cables thru front deck sheet metal openings. Display cable
clips/strain reliefs must be disconnected prior to removing
display assembly (see figure 22).
Figure 34 Removing display assembly
7000 service guide.book Page 98 Thursday, October 16, 2008 12:18 PM