38 7000 Series Oscilloscopes Service Guide
2 Testing Performance
To verify bandwidth
This test checks the bandwidth of the oscilloscope. In this test
you will use a signal generator and a power meter.
1 GHz Models
Test limits at 2 mV/div to 5 V/div
All channels (±3 dB)
dc to 1 GHz
500 MHz Models
Test limits at 2 mV/div to 5 V/div
All channels (±3 dB)
dc to 500 MHz
350 MHz Models
Test limits at 2 mV/div to 5 V/div
All channels (±3 dB)
dc to 350 MHz
100 MHz Models
Test limits at 2 mV/div to 5 V/div
All channels (±3 dB)
dc to 100 MHz
7000 service guide.book Page 38 Thursday, October 16, 2008 12:18 PM