Agilent Technologies 1100 Series Manuel d'instructions Page 147

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 272
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 146
147
Repairing the Pump
Exchanging the Power Supply
Exchanging the Power Supply
When Required:
If defective
Tools required:
Screwdriver Pozidriv #1
Wrench 1/4 inch
Wrench 14 mm
Wrench 7 mm
Wrench 5 mm
Part required:
Power supply 0950-2528
Preparation for this procedure:
Switch off the pump at the main power
switch, remove all cables and tubings
1 Remove the pump from the stack, remove the
front cover, top cover and top foam section
(see Removing the Top Cover and Foam on
page 128).
2 Disconnect all capillaries, tubes and
connectors from the pump head and the
damper. Remove the MCDV cover.
3 Loosen the two Z-panel screws and remove
it.
Pump head
Damper
MCGV
Screws
Z-panel
Vue de la page 146
1 2 ... 142 143 144 145 146 147 148 149 150 151 152 ... 271 272

Commentaires sur ces manuels

Pas de commentaire