Introduction to the Quaternary Pump
Setting the 8-bit Configuration Switch
Forced Cold-Start Settings
Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning the switches 1 and 2 to other positions (other than being
both up) will allow for normal operation.
CAUTION A forced cold start erases all methods and data stored in non-volatile memory.
Exceptions are diagnose and repair logbooks will not be erased.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
To return to normal operation, set switches back to your GPIB or RS 232C
configuration settings.
Stay Resident Settings
Firmware update procedures may require this mode in case of firmware
loading errors.
Table 40 Parity Settings
Switches Parity
7
8
00No Parity
0 1 Odd Parity
1 0 Even Parity
Table 41 Forced Cold Start Settings
Mode Select12345678
TEST/BOOT11000001