Chapter 4: Installing Oscilloscope Measurement Modules
16533/34A Oscilloscope Module (single or multi-card modules)
3 Select the Run icon and the instrument will remind you to connect the
cables. Select OK and wait for the trigger skew calibration to complete.
As trigger skew calibration runs, messages appear in the message box on
screen. When the Ext Trig Skew calibration is complete, the resulting
calibration factors are stored in non-volatile RAM.
4 If a multi-card module is being calibrated, repeat the procedures
beginning with Step 1 on page 72 for each card until all cards have been
calibrated.
5 If all cards have been calibrated, remove the BNC cables from the
instrument.
6 Select Close in the Calibration window.
RUN