To Set the GPIB Address
Each device on the GPIB (IEEE-488) interface must have a unique
address. You can set the multimeter’s address to any value between
0 and 31. The address is set to “
22 when the multimeter is shipped from
the factory. The address is displayed on the front panel when you turn
on the multimeter. See also “
GPIB Address,” on page 91.
1 Turn on the front-panel menu.
A: MEAS MENU
2 Move across to the
I/O MENU choice on this level.
E: I/O MENU
3 Move down a level to the
HP-IB ADDR command.
1: HP-IB ADDR
4 Move down to the “parameter” level to set the address.
Use the left/right and down/up arrow keys to change the address.
22 ADDR
5 Save the change and turn off the menu.
The address is stored in non-volatile memory, and does not change when
power has been off or after a remote interface reset.
<
<
Shift <
On/Off
Auto/Man
ENTER
4
Chapter 4 Remote Interface Reference
To Set the GPIB Address
161