Agilent Technologies AGILENT U2761A Manuel de service Page 18

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 169
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 17
1 Introduction to SCPI
10 U2761A Programmer’s Reference
Using device clear
Device clear is an IEEE- 488 low- level bus message that you can use to
return the instrument to a responsive state (for example, during a lengthy
query).
Different programming languages and IEEE- 488 interface cards provide
access to this capability through their own unique commands. The status
registers, the error queue, and all configuration states are left unchanged
when a device clear message is received.
Device clear performs the following actions:
The instrument’s input and output buffers are cleared.
The instrument is prepared to accept a new command string.
An overlapped command, if any, will be terminated with no “Operation
Complete” indication.
Vue de la page 17