Agilent Technologies AGILENT U2761A Manuel de service Page 44

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 169
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 43
3 APPLy Commands
36 U2761A Programmer’s Reference
APPLy:PULSe
Syntax
APPLy:PULSe [<frequency> [,<amplitude> [,<offset>] ]]
This command outputs a Pulse wave with the specified frequency,
amplitude, and DC offset. The waveform is output as soon as the
command is executed.
Parameters
Example
The following command sets the output frequency to 1.5 kHz, output
amplitude to 3 Vpp, and output offset to 1 V.
APPL:PULS 1500, 3 VPP, 1
Item Type Range of values Default value
frequency Numeric 500 μHz to 5 MHz 1 kHz
amplitude Numeric 40 mVpp to 5 Vpp (Into 50 Ω load)
80 mVpp to 10 Vpp (Into open circuit)
1 Vpp
offset Numeric ±2.48 V (Into 50 Ω load)
±4.96 V (Into open circuit)
0 V
APPL :PULS space frequency , amplitude , offset
Vue de la page 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 168 169

Commentaires sur ces manuels

Pas de commentaire