Chapter 3: Testing Performance
To Test the Time Interval Accuracy
Connect the logic analyzer
1 Using a 6-by-2 test connector, connect channel 0 of Pod 1 to the pulse
generator channel 2 output.
2 Using the SMA cable and the BNC adapter, connect the External Input of
the pulse generator to the Main Signal of the function generator.
Acquire the data
1 Enable the pulse generator channel 2 and trigger outputs (with the LED
off).
2 In the logic analyzer Waveform window, select Run.