Chapter 3: Testing Performance
To Test the Threshold Accuracy
Set up the logic analyzer
1 In the Analyzer Setup window, select the Format tab.
2 Under the Format tab, select Pod Assignment. Unassign the pods that are
assigned to Analyzer 2. To unassign the pods, use the mouse to drag the
pods to the Unassigned Pods column.
Select Close to close the Pod Assignment Window.
3 Under the Format tab, select the Threshold field under Pod 1. Click on the
checkbox next to Apply Threshold Setting to all pods to deselect.