Chapter 6: Replacing Assemblies
To disconnect and connect Mylar flex cables
101
To disconnect and connect Mylar flex cables
Use this procedure when you are instructed to disconnect or connect a Mylar flex cable. Such
cables and their connectors are fragile; mishandling may damage the cable or connector.
To disconnect the cable
1
Pry up the retainer slightly at each end of the connector using a small flat-blade
screwdriver. Do not force the retainer; it should remain attached to the body of the
socket.
2 Gently pull the flex cable out of the connector.
Figure 6-4
Disconnecting a Mylar flex cable
To reconnect the cable
1
Ensure that the cable retainer is up, then insert the ribbon cable into the socket, making
sure to observe polarity of the cable with respect to the connector.
2 Push the ends of the retainer down onto the connector body, using a small flat-blade
screwdriver. The retainer should be flush with the connector body when you are
finished.
Figure 6-5
Connecting Mylar flex sables