Agilent Technologies 81134A Guide de l'utilisateur Page 41

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 92
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 40
Setting Up a Data Pattern Setting Up Generic and Advanced Signals
Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008 41
6 Enable the Channel 1 normal Output:
By pressing the respective softkey in the user interface:
– OR –
By pressing the normal Out softkey for channel 1 next to the
Output port.
The Channel page now looks as follows:
Channel Settings for the
Pattern Signal
To set the channel parameters for channel 2:
1 Switch to Channel 2 by pressing the Ch2 softkey .
2 Switch to the Channel page.
3 Select the pattern mode Data.
4 Select the signal mode NRZ.
5 In the Levels panel, select the predefined level format ECL.
This automatically sets:
–the Amplitude to 800 mV
–the Offset to –1.350 V
–the Term. Voltage to2.0 V
Pattern Mode
PRBS
PRBS Polynome
Signal Mode
ECL Levels
Vue de la page 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 91 92

Commentaires sur ces manuels

Pas de commentaire