Agilent Technologies 81134A Guide de l'utilisateur Page 83

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 92
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 82
Jitter Modulation Using the Agilent 81133A/81134A Pulse Generator
Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008 83
Jitter Modulation
The external source for jitter modulation is applied to the delay
control input. Jitter modulation can be turned on and off
individually for each channel. Either one of two fixed sensitivities
can be selected.
The source for the jitter modulation input is assumed to be either a function
generator or an arbitrary waveform generator. Both have the capability of
setting the output levels. Therefore, the instrument has no capabilities
of adjusting the jitter modulation input.
Saving and Recalling Settings
For permanently saving the instrument setting, nine save/recall
registers are provided.
All parameters are saved, including the data. User interface specifics,
such as the current visible page, are not saved.
When the instrument is turned on, the last used setting is
restored.