Module Specifications (continued)
High Bandwidth, Single Mode Optical/
Electrical Modules 86106B 86109A
OPTICAL CHANNEL SPECIFICATIONS
Optical Channel Unfiltered Bandwidth 28 GHz 30 GHz
Wavelength Range 1000–1600 nm
Calibrated Wavelengths 1310 nm and 1550 nm
Filtered Data Rates (Mb/s) 9953 No internal filters
Sensitivity (smallest average power for mask test) –7 dBm N/A
Transition Time (10% to 90%,
calculated from TR = 0.48/BW optical) 18 ps 16 ps
RMS Noise
Characteristic 10 µW (Filtered) 15 µW (Unfiltered)
18 µW (Unfiltered)
Maximum 15 µW (Filtered) 30 µW (Unfiltered)
30 µW (Unfiltered)
Scale Factor
Minimum 20 µW/division
Maximum 500 µW/division
dc Accuracy (single marker, referenced to ±50 µW ±4% of (reading-channel offset)
average power monitor)
dc Difference (two markers, referenced to average power ±4% of delta
monitor, 50 µW/division)
dc Offset (referenced two divisions from screen bottom) +1 mW to –3 mW
Average Power Monitor (specified operating range) –27 dBm to +3 dBm (2 µW to 2 mW)
Factory Calibrated Accuracy ±5% ±100 nW ±connector uncertainty, 20°C to 30°C
User Calibrated Accuracy ±2% ±100 nW ±power meter uncertainty, <5°C change
Maximum Input Power
Specified 2 mW (+3 dBm)
Maximum Peak 10 mW (+10 dBm)
Fiber Input 9/125 µm, user selectable
Input Return Loss (HMS-10 connector fully filled fiber) 30 dB
1
Specified with 8 point moving average in frequency response
ELECTRICAL CHANNEL SPECIFICATIONS
Electrical Channel Bandwidth 18 and 40 GHz
Transition Time (10% to 90%, 19.5 ps, 18 GHz; 9 ps, 40 GHz
calculated from TR = 0.35/BW)
RMS Noise
Typical 0.25 mV, 18 GHz; 0.5 mV, 40 GHz
Maximum 0.5m V, 18 GHz; 1.0 mV, 40 GHz
Scale Factor
Minimum 1 mV/division
Maximum 100 mV/division
dc Accuracy (single marker, referenced to average ±0.4% of full scale ±2 mV ±1.5% of (reading-channel offset), 18 GHz
power monitor) ±0.4% of full scale ±2 mV ±3% of (reading-channel offset), 40 GHz
dc Difference (two markers, referenced to average ±0.8% of full scale ±1.5% of delta, 18 GHz
power monitor, 50 µW/division) ±0.8% of full scale ±3% of delta, 40 GHz
dc Offset (referenced two divisions from screen bottom) ±500 mV
Input Dynamic Range (relative to channel offset) ±400 mV
Maximum Input Power 16 dBm ±2 Vdc
Nominal Impedance 50 ohm