Specifications
Mainframe Specifications
HORIZONTAL SYSTEM (Time Base)
Scale Factor Full scale is ten divisions.
Minimum 2 ps/div
Maximum 1 s/div
Delay Time offset relative to the front panel trigger input on the instrument mainframe.
Minimum 24 ns (Standard), 28 ns (Option 001)
Maximum 1000 screen diameters or 10 s, whichever is smaller
Time Interval Accuracy 8 ps + 0.1% of reading (dual marker measurement performed at a temperature within ±5°C
of horizontal calibration temperature) 8 ps + 0.1% + 0.5 ps/°C to (5 < absolute temperature
value <15°C)
Time Interval Resolution (screen diameter)/(record length) or 62.5 fs, whichever is larger
Display Units Bits or time
VERTICAL SYSTEM (Channels)
Number of Channels 4 (simultaneous acquisition)
Vertical Resolution 12 bit A/D converter (up to 15 bits with averaging)
Full Resolution Channel Scales Adjusts in a 1-2-5-10 sequence for coarse adjustment or fine adjustment resolution from the
front panel knob
Adjustments Scale, offset, activate filter
Specifications describe warranted performance over the temperature range of +10°C to + 40°C (unless otherwise noted). The specifications are
applicable for the temperature after the instrument is turned on for one (1) hour. This period of continuous operation must occur while self-calibration is
valid. Many performance parameters are enhanced through frequent, simple user calibrations. Characteristics provide useful, non-warranted
information about the functions and performance of the instrument. Characteristics are printed in italic typeface.
Factory Calibration Cycle -For optimum performance, the instrument should have a complete verification of specifications once every twelve (12) months.
General Specifications
This instrument meets Agilent Technologies’ environmental specifications (section 750) for class B-1 products with exception as described for
temperature and condensation. Contact your local field engineer for complete details.
Temperature
Operating 10°C to +40°C (50°F to + 104°F)
Non-operating –40°C to +70°C (–40°F to +158°F)
Humidity
Operating Up to 90% humidity (non-condensing) at +40°C (+104°F)
Non-operating Up to 95% relative humidity at +65°C (+149°F)
Altitude
Operating Up to 4,600 meters (15,000 ft)
Non-operating Up to 15,300 meters (50,000 ft)
Vibration
Operating Random vibration 5–500 Hz, 10 minutes per axis, 2.41 g (rms)
Non-operating Random vibration 5–500 Hz, 10 minutes per axis, 0.3 g (rms); Resonant search, 5–500 Hz
swept sine, 1 octave/min sweep rate, 0.75 g, 5 minute resonant dwell at 4 resonances/raxis
Power Requirements
Voltage 90–132 or 198–264 Vac, 48–66 Hz
Power (including modules) 545 VA; 375 W
Weight
Without modules 12.7 kg (28 lb)
Each Module 1.2 kg (2.6 lb)
Dimensions (excluding handle)
Without front connectors and rear feet 215.1 mm H x 425.5 mm W x 521 mm D (8.47 in x 16.75 in x 20.5 in)
With front connectors and rear feet 215.1 mm H x 425.5 mm W x 584 mm D (8.47 in x 16.75 in x 23.0 in)