Agilent Technologies A.18.00 Manuel d'utilisateur Page 185

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 581
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 184
185
Equivalent Front-Panel Key Commands
HI LIMIT and LO LIMIT
The HI LIMIT and LO LIMIT Data Functions are used to define a measurement
“window” which can be used to detect measured values which are outside the
defined limits. The GPIB commands :HLIMit (high limit) and :LLIMit (low limit)
are used to set these data functions programmatically.
To Turn High and Low Measurement Limit Checking ON and OFF. Use the
:HLIMit:STATe and :LLIMit:STATe commands to turn high and low
measurement limit checking ON and OFF.
Syntax
:HLIMit:STATe ON
:HLIMit:STATe OFF
:LLIMit:STATe ON
:LLIMit:STATe OFF
Example
OUTPUT 714;"MEAS:AFR:DISTN:HLIM:STAT ON"
OUTPUT 714;"MEAS:AFR:DISTN:LLIM:STAT ON"
This turns high and low measurement limit checking ON for the Audio
Frequency Analyzer Distortion measurement.
Vue de la page 184
1 2 ... 180 181 182 183 184 185 186 187 188 189 190 ... 580 581

Commentaires sur ces manuels

Pas de commentaire