Agilent Technologies A.18.00 Manuel d'utilisateur Page 52

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 581
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 51
52
S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb
Chapter 1, Using GPIB
IEEE 488.1 Remote Interface Message Capabilities
Trigger Yes If in remote programming mode and addressed to listen,
the Test Set makes a triggered measurement following the
trigger conditions currently in effect in the instrument.
The Test Set responds equally to the Group Execute Trig-
ger (GET) bus command or the *TRG Common Com-
mand.
GET
MLA
Take Control Yes The Test Set begins to act as the Active Controller on the
bus.
TCT
MTA
Abort Yes The Test Set stops talking and listening IFC
Table 6 Test Set IEEE 488.1 Interface Message Capability (Continued)
Message Type Implemented Response
IEEE
488.1
Message
Vue de la page 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 580 581

Commentaires sur ces manuels

Pas de commentaire