S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb
Chapter 5, Advanced Operations
Status Reporting
Accessing the Hardware Status Register #2 Group’s Registers
The following sections show the syntax and give programming examples (using
the Instrument BASIC programming language) for the STATus commands used to
access the Hardware Status Register #2 Groups registers.
Reading the Condition Register
Syntax
STATus:HARDware2:CONDition?
Example
OUTPUT 714;"STAT:HARD2:COND?"
ENTER 714;Register_value
Reading the Transition Filters
Syntax
STATus:HARDware2:PTRansition?
STATus:HARDware2:NTRansition?
Example
OUTPUT 714;"STAT:HARD2:PTR?"
ENTER 714;Register_value
Writing the Transition Filters
Syntax
STATus:HARDware2:PTRansition <integer>
STATus:HARDware2:NTRansition <integer>
Example
OUTPUT 714;"STAT:HARD2:PTR 256"