Vue de la page 99
3- 22 Keysight B1500A User’s Guide, Edition 14
Installation
Installing Accessories
To install 16495K Fix your necessary 16495 connector plates to the shielding box or something that
will cover the DUT interface. See Keysight 16495 Installation Guide. For the
GSWU installation, the 16495K is required. It is the plate that has the mechanism to
block the light from the cable hole used to pass the cables in the shielding box.
Before starting 1. GSWU is furnished with wires (2 ea.) for the connection between the GSWU
and the DUT interface Guard terminal. If it is too short to connect them, prepare
additional wire.
2. Turn the B1500A off.
Figure 3-5 To Connect SCUU/GSWU
Guard
DUT
To CMU
(slot N)
Force1
Hp
Hc
Sense1
To SMU
(slot N-1)
Force2
Lp
Lc
Sense2
To SMU
(slot N-2)
Force1/CMUH
Sense1
SCUU
Force2/CMUL
Sense2
GSWU