3- 4 Keysight B1500A User’s Guide, Edition 14
Installation
Requirements
Requirements
This section contains information on:
“Power Requirements”
“Operating Environment”
“Storaging/Shipping Environment”
“Installation Requirements”
“Power Cable”
Power Requirements
This instrument can operate from any single-phase AC power source supplying
100-240 V (±10 %) at 50/60 Hz. The maximum power consumption is 900 VA.
Operating Environment
This instrument is specified to operate within the following environmental
conditions:
Temperature: 5 C to 40 C
Humidity: 20 % to 70 % RH, non-condensing
Altitude: 0 to 2,000 m (6,561 ft.)
Storaging/Shipping Environment
This instrument is specified to store/ship within the following environmental
conditions:
Temperature: -20 C to 60 C
Humidity: 10 % to 90 % RH, non-condensing
Altitude: 0 to 4,600 m (15,092 ft.)