Agilent Technologies B1500A Guide de l'utilisateur Page 163

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 204
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 162
Keysight B1500A User’s Guide, Ed ition 14 4- 23
Specifications
Hardware
DC bias function
•DC bias
Range: 0 to ± 25 V
Resolution: 1 mV
Accuracy: ± (0.5 % + 5.0 mV) at the measurement port of the MFCMU or
the MFCMU cable (1.5 m or 3.0 m)
Maximum DC bias current, supplemental characteristics:
Output impedance: 50 , typical
•DC bias monitor
Range: 0 to ± 25 V
Accuracy (open load): ± (0.2 % of reading + 10 mV) at the measurement
port of the MFCMU or the MFCMU cable (1.5 m or 3.0 m)
Sweep characteristics
Available sweep parameters: DC bias voltage, oscillator level, frequency
Sweep type: linear, log
Sweep mode: single, double
Sweep direction: up, down
Number of measurement points: maximum 1001
Impedance range () Maximum DC bias current (A)
50
10
10
-3
100
10
10
-3
300
10
10
-3
1
10
3
1
10
-3
3
10
3
1
10
-3
10
10
3
100
10
-6
30
10
3
100
10
-6
100
10
3
10
10
-6
300
10
3
10
10
-6
Vue de la page 162
1 2 ... 158 159 160 161 162 163 164 165 166 167 168 ... 203 204