114 Chapter 2
Performance Verification Tests
14. Reference Level Accuracy: Agilent E4401B and E4411B
Peak Search (or Search)
Marker, Delta
6. Set the 10 dB step attenuator and analyzer reference level according
to Column 1 and Column 4 of Table 2-22. At each setting, do the
following:
a. Press
Single on the analyzer.
b. Press
Peak Search (or Search).
c. Record the marker delta amplitude reading in Column 5 of
Table 2-22.
d. Add the Actual Attenuator Step to the Analyzer Marker Delta
Amplitude and enter the result in the performance test record.