Agilent E4402B Performance Verification Test Record
Peak-to-Peak Response (3)________ 1.52 dB ±0.19 dB
Option UKB
dc coupled
Band 0
(100 Hz to 3.0 GHz)
Maximum Response (1)________ 1.0 dB
±0.19 dB
Minimum Response –1.0 dB (2)________
±0.19 dB
Peak-to-Peak Response (3)________ 2.0 dB
±0.19 dB
Non-Option UKB
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39441006 or greater)
Maximum Response (4)________ 0.76 dB
±0.11 dB
Minimum Response –0.76 dB (5)________
±0.11 dB
Peak-to-Peak Response (6)________ 1.52 dB
±0.11 dB
Non-Option UKB
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39441006 or greater)
Maximum Response (7)________ 0.76 dB
±0.11 dB
Minimum Response –0.76 dB (8)________
±0.11 dB
Peak-to-Peak Response (9)________ 1.52 dB ±0.11 dB
Table 3-4 Agilent E4402B Performance Verification Test Record