Agilent Technologies E1401-90493 Manuel d'utilisateur Page 469

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 642
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 468
Chapter 3 469
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
Minimum Response –1.5 dB (2)________ ±0.28 dB
Peak-to-Peak Response (3)________ 3.0 dB
±0.28 dB
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39441006 or greater)
Maximum Response (4)________ 1.5 dB
±0.28 dB
Minimum Response –1.5 dB (5)________
±0.28 dB
Peak-to-Peak Response (6)________ 3.0 dB
±0.28 dB
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39441006 or greater)
Maximum Response (7)________ 1.5 dB
±0.28 dB
Minimum Response –1.5 dB (8)________
±0.28 dB
Peak-to-Peak Response (9)________ 3.0 dB
±0.28 dB
0 to 55
° C
Band 0
(1 MHz to 3.0 GHz)
Maximum Response (1)________ 2.0 dB
±0.28 dB
Minimum Response –2.0 dB (2)________ ±0.28 dB
Peak-to-Peak Response (3)________ 4.0 dB
±0.28 dB
Table 3-4 Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty
Vue de la page 468
1 2 ... 464 465 466 467 468 469 470 471 472 473 474 ... 641 642

Commentaires sur ces manuels

Pas de commentaire