Agilent Technologies E1401-90493 Manuel d'utilisateur Page 562

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 642
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 561
562 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
(Serial Prefix US4510,
SG4510 or MY4510,
Option 1DR)
10 kHz (1)________ –98 dBc/Hz
±2.44 dB
20 kHz (2)________ –104 dBc/Hz
±2.44 dB
30 kHz (3)________ –110 dBc/Hz ±2.44 dB
(Serial Prefix
US4510,
SG4510 or MY4510,
Non-Option 1DR)
10 kHz (1)________ –90 dBc/Hz
±2.44 dB
20 kHz (2)________ –104 dBc/Hz
±2.44 dB
30 kHz (3)________ –110 dBc/Hz
±2.44 dB
8. Noise Sidebands - Wide
Offsets
Note: Enter data in the appropriate section below depending upon
whether or not Option 120 (ACPR Dynamic Range Extension) is
installed.
Non-Option 120
Offset from 1 GHz signal
100 kHz (1)________ –118 dBc/Hz
±2.44 dB
1 MHz (2)________ –125 dBc/Hz ±2.44 dB
5 MHz (3)________ –127 dBc/Hz
±2.44 dB
10 MHz (4)________ –131 dBc/Hz
±2.44 dB
Option 120
Offset from 1 GHz signal
100 kHz (1)________ –118 dBc/Hz
±2.44 dB
1 MHz (2)________ –133 dBc/Hz ±2.44 dB
5 MHz (3)________ –135 dBc/Hz
±2.44 dB
10 MHz (4)________ –137 dBc/Hz
±2.44 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty
Vue de la page 561
1 2 ... 557 558 559 560 561 562 563 564 565 566 567 ... 641 642

Commentaires sur ces manuels

Pas de commentaire