Probing 2
To Change an N2825A Head’s ID Resistor
N2820/1A User’s Guide 47
8 On Infiniium oscilloscopes,
a Click the Setup > Probe Configuration menu command.
b Select the tab for the probe’s primary channel. N2821A
probes only have a primary channel.
c In the User Def Head Resistance field, enter the resistance of
the DUT’s R
SENSE
resistor in ohms.
d In the dialog box, click Add Head to give your head with
associated external R
SENSE
resistor a unique name.
9 On InfiniiVision oscilloscopes,
a Press the front- panel channel key that is associated with
the probe’s primary input.
b Click the Probe softkey.
c Press the R-Sense softkey and enter the resistance of the
R
SENSE
resistor that you are using.
10 For future reference, mark your different N2825A heads
using a marker pen or tape and record the information in
Table 6 on page 48. In the tables fourth column, record the
name that you entered in the Probe Configuration dialog
box.