Agilent Technologies N1022A Guide de l'utilisateur Page 56

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 82
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 55
56 N2820/1A User’s Guide
3 Characteristics and Specifications
Zoomed-In Input Voltage Range Versus Supply Voltage
Zoomed-In Input Voltage Range Versus Supply Voltage
For the zoomed- in probe input, the following graph shows
the typical input voltage range allowed based on the supply
voltage that you are measuring. For example, when
measuring a 5V input, the input voltage range is \45 mV.
Figure 31 Zoomed-In Input Voltage Range Versus Supply Voltage
NOTE This plot can be used with N2825A heads (user-defined R
SENSE
) to verify
current range versus supply voltage.
Vue de la page 55