Agilent Technologies Sprinkler 54753a Manuel d'utilisateur Page 14

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 44
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 13
Display Commands
Command
:DISPlay:ASSign[?] {CHANnel<number> |
FUNCtion<number> | RESPonse<number> |
WMEMory<number> | FFT},{UPPer | LOWer}
Assigns the specified channel, waveform, function, TDR response, or FFT to
a particular portion of the waveform area on the screen. This command has
no effect when the graticule format is single.
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
12